ATK-3 & ATK-4 | Accuracy Testing Kit

Accuracy Testing Kit for calibration of the meter testers in-house

ATK-3 et ATK-4 | Accuracy Testing Kit | Calibration of Meter Device | Metering

ATK-3 & ATK-4 (Accuracy Testing Kit): To guarantee optimal working conditions, it is strongly advise to calibrate all test equipment at least once a year or as per local regulations.

An available accuracy test kit makes it possible to perform an accuracy validation on a testers in-house. The materials needed for calibrating the units are:

  • An isolated power supply, 120V–240V rated at least 150 VA
  • A reference standard traceable to NIST with minimum accuracy of 0.05%
  • A Probewell accuracy testing kit ATK-3 (for three-phase models)
  • A Probewell accuracy testing kit ATK-4 (for single-phase models)

ATK-4 SKU: 9910-0290

ATK-3 SKU: 9910-0285

ATK-3 & ATK-4 (Accuracy Testing Kit): To guarantee optimal working conditions, it is strongly advise to calibrate all test equipment at least once a year or as per local regulations.

An available accuracy test kit makes it possible to perform an accuracy validation on a testers in-house. The materials needed for calibrating the units are:

  • An isolated power supply, 120V–240V rated at least 150 VA
  • A reference standard traceable to NIST with minimum accuracy of 0.05%
  • A Probewell accuracy testing kit ATK-3 (for three-phase models)
  • A Probewell accuracy testing kit ATK-4 (for single-phase models)

ATK-4 SKU: 9910-0290

ATK-3 SKU: 9910-0285

ATK-3 et ATK-4 | Accuracy Testing Kit | Calibration of Meter Device | Metering


ATK-3 & ATK-4 | Accuracy Testing Kit | Installation

ATK-3 | Accuracy Testing Kit | Calibration of Meter Device | Metering
ATK-3 | Accuracy Testing Kit | Calibration of Meter Device | Metering

The accuracy of MT-1/NT4, MT-1/NT5, MT-1/NT9, MT-1/WT1 and MT-1/WT3 meter tester can be checked against a NIST-traceable standard using the following test procedure. The accuracy is tested on up to 17 different current test points at both 1 (0°) and 0.5 lag (60°) power factors in Wh.

Additionally the MT-1/NT9 and MT-1/WT3 does VARh at 30° and 90°. The accuracy test can be performed using a stabilized and isolated AC power source at either 120 or 240 volts.

Accuracy Testing Kit | ATK-3: MT-1/WT3 and MT-1/NT9 | ATK-4: MT-1/WT1, MT-1/NT4 and MT-1/NT5 | Meter Tester Accuracy | Metering

The accuracy of  MT-1/NT4, MT-1/NT5, MT-1/NT9, MT-1/WT1 and MT-1/WT3 meter tester can be checked against a NIST-traceable standard using the following test procedure. The accuracy is tested on up to 17 different current test points at both 1 (0°) and 0.5 lag (60°) power factors in Wh.

Additionally the MT-1/NT9 and MT-1/WT3 does VARh at 30° and 90°. The accuracy test can be performed using a stabilized and isolated AC power source at either 120, 240 or 480 volts.

Accuracy Testing Kit | ATK-3: MT-1/WT3 and MT-1/NT9 | ATK-4: MT-1/WT1, MT-1/NT4 and MT-1/NT5 | Meter Tester Accuracy | Metering